Praetz S, Grötzsch D, Schlesiger C, Motz D, Würth M, Zimmermann R, Lucka R, Malzer W, Lützenkirchen-Hecht D, Renz F, Kanngießer B (2025)
Publication Type: Journal article
Publication year: 2025
Book Volume: 96
Article Number: 035120
Journal Issue: 3
DOI: 10.1063/5.0253653
For the investigation of spin crossover (SCO) complexes by x-ray absorption spectroscopy (XAS) measurements in the tender to hard x-ray regime, a transmission heating cell has been developed. Some SCO complexes show a spin transition at certain temperatures above room temperature (RT). To measure a reliable hysteresis curve, high temperature stability (preferably <0.5 °C) as well as minimal overheating of the setpoint temperatures (<1 °C) has been achieved. The cell is designed for use in our von Hámos XAS laboratory spectrometer. The SCO complex [Fe(bzimpy)
APA:
Praetz, S., Grötzsch, D., Schlesiger, C., Motz, D., Würth, M., Zimmermann, R.,... Kanngießer, B. (2025). In situ heating cell for temperature dependent transmission x-ray absorption spectroscopy (XAS) measurement with a laboratory based spectrometer. Review of Scientific Instruments, 96(3). https://doi.org/10.1063/5.0253653
MLA:
Praetz, S., et al. "In situ heating cell for temperature dependent transmission x-ray absorption spectroscopy (XAS) measurement with a laboratory based spectrometer." Review of Scientific Instruments 96.3 (2025).
BibTeX: Download